Merilyzer ClotQuant 2

Coagulation Analyzer

Machines & equipment

Meril Merilyzer ClotQuant 2 semi auto coagulometer is designed for PT and APTT analysis, supporting routine coagulation testing in labs.

01 / A CLOSER LOOK

Features that
define the product.

  • 2 Testing channels, 16 Sample pre-warming positions and 4 Reagent pre-warming positions
  • ClotQuant 2 is equipped with Electronically-linked pipette to reduce manual errors, built-in thermal printer and on-board stop watch
  • In Sample run, high and low flags will be displayed and printed in run receipt, based on reference range entered by user
  • In QC run, based on mean and SD values entered by user, SD flags (+1, +2, +3,-1, -2,-3), over and under will be displayed and printed in QC run receipt
  • QC Levey Jennings to track performance of controls (Low, Medium and High) for 30 days. 40 tests can be stored for each level of control i.e. total 120 results
  • ClotQuant 2 ensures accurate results by monitoring the exact Sample Incubation Time. Temperature does not exceed above or below the user defined limits
  • ClotQuant 2 detector channels can be aligned in any positions to ensure operators convenience

02 / HOW IT WORKS

The technology.

Measuring Principle - Advanced scattered light

03 / TECHNICAL DETAILS

Specifications.

Manufacturer-listed information
CategorySpecification
Measuring PrincipleAdvanced scattered light
Reagent TypeOpen
Reagent Consumption< 50 μL
Tests PerformedPT, APTT, TT, Fibrinogen, and clotting factors
Test Wavelength470 nm
Testing Channels2 channel (ClotQuant 2), 4 channel (ClotQuant 4)
Sample Pre-warming Positions16 positions (ClotQuant 2), 24 positions (ClotQuant 4)
Reagent Pre-warming Positions4 positions (ClotQuant 2), 6 positions (ClotQuant 4)
Dispense PrecisionCV < 2%
Temperature Control Precision37 ± 0.3°C
RepeatabilityActivated partial thromboplastin time tolerance ± 2 s
StabilityActivated partial thromboplastin time tolerance within one hour ± 2 s
Channel ConsistencyEach channel activated partial thromboplastin time tolerance ± 2 s
Linearity ErrorFibrinogen concentration linear correlation r² > 0.98
Memory5000 test results (500 Patient ID, 10 tests for each ID)
Power Supply230 VAC ± 10%, 50Hz
Communication InterfaceRS232 interface, PC connecting
PrinterBuilt-in thermal printer, 57 mm paper width
ReportDate and ID wise
Work Environment15°C - 30°C
Humidity≤ 90%
Dimensions (L × W × H)400 mm × 274 mm × 350 mm
WeightApprox. 10 kgs